Biobase China Nucleic Acid Extractor 1-96 Samples Magnetic Bead Automatic DNA Rna Extraction System

Still deciding? Get samples of $ !
Request Sample

Product Description

i

Basic Specifications

Model NO.
BNP32
Product Name
Nucleic Acid Purification Extraction System
Sample Quantity
32
Processing Volume
60µl-1000µl
Sample Volume
20-500µl
Reagent Type
Magnetic Bead Open Platform
Extraction Time
8-60 Min/Round
Production Capacity
20000 Box/Month

Product Highlights

7-inch touch screen, easy to use, fast response.
User-defined cracking and elution temperature.
UV disinfection function, time range 1min-24hour.
Automatic control system, no need to connect to a computer.
Free programming to meet the needs of different reagents.
Open system, fully automatic, stable results and good repeatability.
Extract rapidly within 15-40 minutes; 32/48 samples simultaneously.
📷

Detailed Visuals

Technical Parameters

Magnetic Bead Recovery>98%
Extracting Difference Between HolesCV≤3%
Heating Temperature8 independent heating modules, customizable lysis and elution
Oscillating MixingLow, medium, and high gears adjustable
Internal ProgramStores >100 groups of programs
Program ManagementUSB disk import/export capabilities
Safety Door DesignAutomatic suspension when opened
Ultraviolet IrradiationYes
Gross Weight (kg)51
Packing Size(W*D*H)mm580*510*700
?

Frequently Asked Questions

How many samples can the BNP32 process at once?
The system is designed to handle between 1 to 32 samples per round, making it ideal for various laboratory throughput requirements.
What is the typical extraction time for a cycle?
The extraction process is very rapid, typically taking between 15 to 40 minutes per round depending on the specific reagent used.
Is the system compatible with third-party reagents?
Yes, it features a Magnetic Bead Open Platform, which allows for free programming to meet the needs of different reagent kits.
Does the machine require a computer connection?
No, the unit features an automatic control system with a built-in 7-inch touch screen, allowing for completely standalone operation.
How is contamination prevented during operation?
The system includes a built-in UV disinfection function and a safety door design that automatically pauses the program if opened, ensuring a sterile and safe environment.
Can the temperature be adjusted for different protocols?
Yes, there are 8 independent heating modules that allow users to customize lysis and elution temperatures according to their specific laboratory protocols.